SS24-316L, SS36-316L Filter Housing Series
1)Installing the Filter Housing (SS24-316L and SS36-316L Filter Housing Series)
Locate the Inlet (marked as “IN” on the fl ange) and Outlet (marked as “out” on the fl ange) connections on the housing cylinder. Connect the
water supply line piping to the Inlet connection and the fi ltered water piping serving the building to the Outlet connection. Shut off valves
are to be installed on each side of the housing (see Note). DO NOT install the fi lter housing backwards by connecting the water supply line
piping to the Outlet. The drain plugs should then be removed and either a drain line assembly (with two valves) or two hose bibs (faucets)
should be installed. See Figure 1 for proper installation. Install fi lter cartridges per Step 2 prior to fi lter housing start-up.
Note: When the housing is installed in close proximity to the water meter or pressure tank, the main shut off valve may be used for the
Inlet water control valve.
Note: Installation must comply with state and local plumbing codes.
SLUDGE CHAMBER
DRAIN 3/4" NPT
G DIA. 3 HOLES
EQUALLY SPACED ON H
DIA. BOLT CIRCLE
VENT PLUG - 1/4" NPT
LOCK NUT
COVER
PRESSURE PLATE
ASSEMBLY
COVER GASKET
V-BAND CLAMP
SPRING AND SEAL PLATE
ASSEMBLY
CARTRIDGE POSTS
SLUDGE CHAMBER
DRAIN PLUG 3/4" NPT
MOUNTING LEGS
CLEAN CHAMBER
DRAIN PLUG 3/4" NPT
CARTRIDGE
PEDESTALS
FLOW BAFFLE
1
2
6
7
8
9
10
11
12
15
12
Plan View
Elevation View
(See Plan View for True Orientation)
F
INLET OUTLET
T-BOLT
KIT
INLET
B
A
C
HEIGHT REQUIRED TO
CHANGE CARTRIDGES
5
4
14
SLEEVE
RETAINING RING
CENTER ROD
OUTLET
DE
WASHER
3
6
Page view 5
1 2 3 4 5 6 7 8

Comments to this Manuals

No comments